There is no specific OSHA requirement for fume hood monitors.  The applicable reference is the OSHA Laboratory Standard, which requires the chemical hygiene plan to include:

1910.1450(e)(3)(iii)

A requirement that fume hoods and other protective equipment are functioning properly and specific measures that shall be taken to ensure proper and adequate performance of such equipment.

 

While newer laboratory hoods generally include a built-in flow monitoring device, older hoods are commonly tested periodically using a portable flow monitoring device such as a vaneometer.   Testing frequency should be addressed in your chemical hygiene plan; many facilities perform such testing quarterly or at least semi-annually.  Additional local codes or insurance requirements may make more frequent testing necessary.

 

Russ Phifer

 

 

Russ Phifer

WC Environmental, LLC

1085C Andrew Drive

West Chester, PA  19380

610-696-9220x12/ fax 610-344-7519

rphifer**At_Symbol_Here**wcenvironmental.com

 

P Please consider your environmental responsibility before printing this e-mail or any other document

 

From: DCHAS-L Discussion List [mailto:DCHAS-L**At_Symbol_Here**LIST.UVM.EDU] On Behalf Of Hilbrandt, Robert
Sent: Thursday, July 16, 2009 11:04 AM
To: DCHAS-L**At_Symbol_Here**LIST.UVM.EDU
Subject: [DCHAS-L] Cite Reference

Could you site the requirement for fume hood monitors.  We are located in New York State (upstate) and are not aware of such a requirement.

Robert W. Hilbrandt Jr., RMS
Radiochemistry Technical Director / QA Manager
Westchester County Dept. Laboratories & Research
10 Dana Road, Valhalla, NY  10595
Tel: 914-231-1531  Fax: 914-231-4458
rwh2**At_Symbol_Here**westchestergov.com
Visit us at http://blog.w estchestergov.com/index.php?option=com_content&task=view&id= 492&Itemid=1009

Checked by AVG - www.avg.com
Version: 8.5.387 / Virus Database: 270.13.16/2241 - Release Date: 07/16/09 05:58:00

Previous post   |  Top of Page   |   Next post