Demystify:
I agree with Rob. Osmium tetroxide has a TLV TWA of 0.0002 ppm or 0.0016 mg/m3. You have no choice but to use air-supplied. And I just looked at 3M cartridge information and Supplied Air is their recommendation. And since OS4 still a solid at room temperature (MP 104 F - 40 o C) there really could be stuff lurking there depending on conditions.
-----Original Message-----
From: ILPI Support <info**At_Symbol_Here**ILPI.COM>
To: DCHAS-L <DCHAS-L**At_Symbol_Here**MED.CORNELL.EDU>
Sent: Mon, Aug 5, 2013 5:20 pm
Subject: Re: [DCHAS-L] demolition of lab with osmium tetroxide contamination
Not an expert at that, but for the hood I'd go with a full face supplied air respirator. While it's unlikely there is anything volatile remaining, OsO4 attacks the corneas and I would not rely on a cartridge which probably has not been tested against Os4.
I'd like to know if anyone has dealt with the demolition of a lab and removal of a fume hood in which osmium tetroxide was used? Other than recommending respiratory protection (N100 or P100 cartridges), nitrile gloves and coveralls during tear-out, are there any other precautions we should keep in mind?
The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary**At_Symbol_Here**dchas.org. The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.