From: "Osprey, James" <ospreyj**At_Symbol_Here**NOVATECH.CA>
Subject: Re: [DCHAS-L] Flashback prevention
Date: Tue, 19 Nov 2019 17:49:46 +0000
Reply-To: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>
Message-ID: 340C8DA756B86A49BA784F71DF66CAFB01550E6B7A**At_Symbol_Here**Mail.novatech.local
In-Reply-To


My experience is that flame arrestors are frequently used with on-line analyzers that are connected to a process for continuous on-line monitoring. This need does depend on the analyzer technology and the analyzer manufacturers approvals. When required, the arrestors are fitted at the sample inlet to the analyzer. With a GC, there is not commonly a direct connection to the process from the analyzer, being isolated by the injection valve. In the event, the injection flows are very small (5-100ë1Ú4l per measurement cycle) and dead volume on the sample inlet side is extremely critical.

The use of hydrogen with GC‰??s is usually as carrier gas but can also be as fuel gas, in either case we are talking about very high purity hydrogen. I have never seen flame arrestors in these lines between the H2 cylinders and the analyzer. Where analyzers vent to a hazardous area, it is not uncommon to have a flame arrestor in the vent header, however these do give ongoing maintenance and are frequently removed (post installation)

There is a catch 22 with these devices, those that are tested and approved to a standard, invariably are tested with an explosive mixture of the flammable in air. As example, the flame transmission test for group C, D gases is usually done with a group B gas. In principle, It, should be possible to carry out tests for a group B gases with a group B gas supercharged with Acetylene but I have not come across such as commercially available.

The other insurmountable problem is, that those standards that are available are for large devices to be fitted on much larger line sizes than those to be encountered in normal analyzer sample lines. As such these commercially available arrestors tend to be large, much too large for small analyzer flows. In consequence, when required, the analyzer manufacturer supplies dedicated devices for the purpose - I am not aware that these meet or are designed to a any kind of recognized international standard.


James Osprey, C. Phys, C. Eng.
Technical Advisor

Int +44 759 412 4934
ospreyj**At_Symbol_Here**novatech.ca

From: ACS Division of Chemical Health and Safety On Behalf Of James Saccardo
Sent: November-19-19 5:00 PM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: Re: [DCHAS-L] Flashback prevention

Steve and Rich,
Irrespective of what the law or code states, we must ask what is reasonable and prudent. It does seem reasonable to put them on the AA, especially since the cost is inexpensive. And if you‰??re going to put them on one flammable gas, you might as well put them on all. The cost is minor and it is a good exercise in safety, and evaluate the piping system. At a time when safety buy-in is meager at best, and in need of a paradigm shift, safety professionals must stand of the side of reasonable and prudent, and then consider ‰??are we in compliance‰??. Nothing wrong with exceeding the requirements of the code. You will never measure the incidents that are prevented, but they will always measure the incidents that occur on your watch.

Consequently, what concentration of hydrogen are you using on your GC?
James

From: ACS Division of Chemical Health and Safety On Behalf Of Richard Palluzi
Sent: Tuesday, November 19, 2019 5:45 AM
To: mailto:DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: Re: [DCHAS-L] Flashback prevention

These are only mandatory on welding cylinders. They are designed to protect against a flashback if the hoses fail

Sent from my Verizon Wireless 4G LTE Droid
On Nov 18, 2019 4:59 PM, Steven Smith wrote:
Hi All,

One of the instrumentation labs I work in recently had a report suggesting that both the hydrogen and acetylene tanks be equipped with flashback adapters. These are for tanks used for a GC and a Flame AA. The manufacturer, however, suggests that these regulators aren‰??t necessary. I was wondering if anyone on this listserv could comment on whether they use flashback adapters for tanks attached to GC‰??s and FAA instruments.

Thanks,

Steve

Dept of Chemistry
Memorial University
--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at mailto:membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas
--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at mailto:membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas
--- For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at mailto:membership**At_Symbol_Here**dchas.org Follow us on Twitter **At_Symbol_Here**acsdchas

Confidentiality Warning:
This message and any attachments are intended only for the use of the intended recipient(s), are confidential, and may contain privileged information. If you are not the intended recipient, you are hereby notified that any review, retransmission, conversion to hard copy, copying, circulation or other use of this message and any attachments is strictly prohibited. If you are not the intended recipient, please notify the sender immediately by return e-mail, and delete this message and any attachments from your system. Thank you.
Information confidentielle:
Le prÌ©sent message, ainsi que tout fichier qui pourrait y Ì»tre joint, sont envoyÌ©s Ì l'intention exclusive de son ou de ses destinataires; ils sont de nature confidentielle et peuvent constituer une information privilÌ©giÌ©e. Nous avertissons toute personne autre que le destinataire prÌ©vu que tout examen, rÌ©acheminement, impression, copie, distribution ou autre utilisation de ce message et de tout fichier qui y est joint est strictement interdit. Si vous n'Ì»tes pas le destinataire prÌ©vu, veuillez en aviser immÌ©diatement l'expÌ©diteur par retour de courriel et supprimer ce message et tout document joint de votre syst̬me. Merci.

---
For more information about the DCHAS-L e-mail list, contact the Divisional membership chair at membership**At_Symbol_Here**dchas.org
Follow us on Twitter **At_Symbol_Here**acsdchas

Previous post   |  Top of Page   |   Next post



The content of this page reflects the personal opinion(s) of the author(s) only, not the American Chemical Society, ILPI, Safety Emporium, or any other party. Use of any information on this page is at the reader's own risk. Unauthorized reproduction of these materials is prohibited. Send questions/comments about the archive to secretary@dchas.org.
The maintenance and hosting of the DCHAS-L archive is provided through the generous support of Safety Emporium.