Subject: [DCHAS-L] CHAS Symposium Highlight: From Laboratory Bench to Community

Date: Aug 14, 2026 16:41 UTC

Author: Chung, Amanda Brittany <0000209df1e25ae8-dmarc-request**At_Symbol_Here**LISTS.PRINCETON.EDU>

Next by Date

Subject: [DCHAS-L] Last call for web site feedback

Date: Aug 18, 2026 18:33 UTC

Author: Marta Guron <marta.guron**At_Symbol_Here**GMAIL.COM>

Thread context

From: Charlie Gates <charlie**At_Symbol_Here**GATES.HOUSE>

Subject: Re: [DCHAS-L] Gas Sensor

Date: Aug 14, 2026 19:26 UTC

Reply-To: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>

In-Reply-To: Re: [DCHAS-L] Gas Sensor

Demystify: 
Dear Ruth:

In my old lab at UCSF, we had a room exclusively dedicated to LN2 storage that remained unoccupied. That room had an oxygen sensor (an MSA X5000 with an O2 sensor) which was wired to a strobe light on the outside of the door. I have since moved to a lab in an older building that lacks this installed equipment, though the spaces housing the LN2 dewars are reasonably well ventilated and are occupied.

You may want to talk to your on-site industrial hygienist or safety personnel to see if they have any more specific guidance.

I hope you all have a nice weekend,

Charlie


On Tue, Aug 11, 2026 at 6:45 AM James Saccardo <00002534240fdb40-dmarc-request**At_Symbol_Here**lists.princeton.edu> wrote:

Hi Ruth,

For natural gas, the sensor is built right into the human olfactory gland. There are mercaptans in the gas that will aid any human in detecting a leak at the ppm level. Action must be taken immediately, humans can quickly become acclimated.

 

For LN2 it depends on the volume, a few liters should no be a big deal, now if you have 100L or more, and a leak can displace all breathable oxygen, then you might want to consider having a hardwired continuous oxygen monitor installed. Check with you local fire official, they may provide guidance for cryogens and what volume you would need O2 sensors.

 

Implementing a good standard operating procedure may be your best practice.

Best,

James

 

 

From: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU> On Behalf Of Murphy, Dr. Ruth Ann
Sent: Wednesday, August 5, 2026 5:19 PM
To: DCHAS-L**At_Symbol_Here**PRINCETON.EDU
Subject: [DCHAS-L] Gas Sensor

 

* External Sender - Proceed Cautiously with Links and Attachments. *

Good Afternoon,

 

Are gas sensors required in chemistry labs with natural gas utilities and/or liquid nitrogen use?  If so, what type of sensor is needed?  Thank you!

 

Best,

 

Ruth Ann Cook Murphy, PhD

Professor of Chemistry

University of Mary Hardin-Baylor

900 College Street

Belton, TX 76513

(254) 295-4542

 